TXF1270 X-Ray HV Power Supply

TXF1270

Teslaman TXF1270 series X-ray high voltage power supply is a compact power supply with negative, positive, or bipolar output options, with a power range from 1.8 kW to 4.5 kW, and an output voltage range of 160 kV to 450 kV, suitable for various application needs. The power supply adopts an active power factor correction circuit, with a high power factor and low input current, and reduces line-related electromagnetic interference to a minimum. TXF1270 adopts a proprietary inverter topology technology to make the power supply more efficient and higher in power density. The high-voltage part of the power supply is packaged in a solid-state way to reduce volume and weight and improve the reliability and maintenance convenience of the product. The power supply uses DSP technology based on SMT control circuits and provides various interfaces such as USB, RS-232, Ethernet, and analog signals to facilitate integration into OEM systems. The power supply has two DC outputs, and the stable emission current regulation circuit controls the filaments power supply, providing precise and stable current to the X-ray tube. In addition, the power supply also has a comprehensive fault diagnosis circuit, arc monitoring, arc extinguishing, and arc counting functions.

Typical Application:Non-destructive testing,X-ray scanning,Secure applications,Medical applications,Al Visual Recognition.

Reference Back to list

Features

  • 160kV,225kV,320kVand450kV Model
  • 1.8kW,3kW,4kW,4.5kW output power available
  • Build in PFC circuit
  • Integrated double filament power supply
  • Digital interface USB
  • Ethernet and RS-232
  • Solid Encapsulation

CONTACT US

Address:16 Renxian Street, High-tech Industrial Zone, Dalian, China
Telephone:+86411-84754722
Mobile:+8615642371743
Wechat:15642371743
Email:chengxingqi@teslamanhv.com
URL:www.teslamanhv.com
DaLian Teslaman XingQiCheng

Product Specifications


Input Voltage

AC220V±10%, 50/60Hz, PFC input≥0.98.

Input Current

<30A.

Output Voltage

Accuracy

0.25%.

Stability

≤ 0.1% per 8 hours, after 1 hour warm up.

Load

±0.05% of rated output voltage for a full load change.

Line

±0.05% of rated output voltage over specified input voltage range.

Temperature Coefficient

50ppm/°C.

Emission Current

Accuracy

0.25%.

Temperature coefficient

100ppm/°C.

Load

±0.05% of rated output current for a change from 30% to 100% of rated output voltage.

Line

±0.05% of rated output current over specified input voltage range.

Filament

Output

0-6 amps at a compliance of 10Vdc, maximum.

Dual Focal Spot

Small and large, selectable via interface signal.

Configuration

DC filament drive. Closed loop emission control regulates filament setting to provide desired X-Ray tube emission current.

Polarity

Specify negative or positive when ordering.

Interface

Remote interface

Analog、USB、Ethernet and RS-232.

Software

Software for control and power supply information.

Environment

Work

0°C To +50°C.

Storage

-40°C To +85°C.

Mains Input Connector

97-3102A-24-11P.

Interface Connector

Digital—Ethernet, RS-232 and USB. Analog—25 pin connector.

Cooling

Forced air.