TXF1250
Teslaman TXF1250 series high- voltage power supply is an X-ray generator module specially designed for OEM applications, with a voltage output of up to 160kV and a power of 1200W. The power supply has a universal input, a small package size and three standard digital interfaces, which can be more simplified and integrated into your X-ray analysis system. The whole series of models can be used in any X-ray tube design of suspended filament (negative high voltage polarity) or grounding filament (positive high voltage polarity). The control circuit based on DSP enables the power supply to provide excellent emission current regulation and excellent stability.
Typical Application:Blood irradiation, nondestructive testing, crystal testing,electroplating measurement,diamond testing, mineral analysis,X-ray fluorescence, X-ray diffraction, plastic sorting.
Reference Back to list
Features
- Compact and light weight.
- Optional parameters-20kV to - 160kV 300W 600W and 1200W.
- Universal input and power factor correction
- Support cold cathode or hot cathode X-ray tubes.
- Standard digital interfaces:USB Ethernet and RS-232.
- User programmable and arc detection functions.
CONTACT US
Telephone:+86411-84754722
Mobile:+8615642371743
Wechat:15642371743
Email:chengxingqi@teslamanhv.com
URL:www.teslamanhv.com
Product Specifications
Input voltage
Input power factor correction AC100-240V ±10%;300W power supply, 47-63Hz, 4.6A.
AC200-240V ±10%
600W power supply, 47-63Hz, 4.3A. 1200W power supply, 47-63Hz, 8.2A.
Output voltage
8Models - 20kV, 30kV, 40kV, 50kV, 60kV, 70kV, 75kV, 100kVAnd 160kV.
Output polarity
Negative polarity - used for suspended filament X-ray tube Positive polarity - used for grounding filament X-ray tubes
Power
300W, 600W, 1200W. Customizable
Output voltage regulation
Within the specified input voltage range, ≤ 0.01% of the rated output voltage. Full load change, ≤ 0.01% of the rated output voltage.
Emission current regulation
Within the specified input voltage range, ≤ 0.01% of the rated output current. When the rated output voltage changes from 30% to 100%, ≤ 0.01% of the rated output current. When kV < 30% of the full range output, the filament is disabled.
Ripple
Under rated output conditions, it is better than 1% rms (0.1% rms optional).
Stability
After 2 hours of startup, it is better than 25ppm/hour.
Temperature coefficient
Voltage and current are better than50ppm/℃.
Environment
Temperature range
Operating temperature
0 C to 40 C
Storage temperature
-40 C to 85 C
Humidity
20% to 85% relative humidity, no condensation.
Filament configuration
Closed-loop emission control adjusts the filament setting to provide the desired X-ray tube emission current.
Two types are available
suspension filament (AC output is referenced by negative output voltage) and grounding filament (DC output is referenced by ground). Output
Control interface
Local interface
Adjust the filament limit and preheating level through the potentiometer.
Remote interface
Standard USB, Ethernet and RS232. All digital monitoring is a 2% accuracy specification.
Control software
Provide an example of Windows graphical user interface.
High voltage enable
Hardware-based, dry contact closure will enable the power supply to enter high- voltage mode.
Monitoring signal
The voltage and current monitoring signal are proportional, 0-10VDC is equal to 0-100% full range, and the accuracy is 1%.
Cooling
Forced ventilation.
External dimensions
300/600W
120.65mm high, 152.4mm wide and 304.8mm deep.
1200W
120.65mm high, 304.8mm wide and 304.8mm deep.
160kV
226.7mm high, 482.6mm wide and 546.1mm deep.
Weight
300/600W
7.5Kilogram
1200W
15Kilogram
160kV
70Kilogram
Input power connector
IEC320 with EMI filter
Output connector