TXF1272 X-Ray HV Power Supply

TXF1272

Teslaman TXF1272 series X-ray high voltage power supply is a compact power supply with negative, positive, or bipolar output options, with a power of 6.0 kW, and an output voltage range of 160 kV to 450 kV, suitable for various application needs. The power supply adopts an active power factor correction circuit, with a high power factor and low input current, and reduces line-related electromagnetic interference to a minimum. TXF1272 adopts a proprietary inverter topology technology to make the power supply more efficient and higher in power density. The high-voltage part of the power supply is packaged in a solid-state way to reduce volume and weight and improve the reliability and maintenance convenience of the product. The power supply uses DSP technology based on SMT control circuits and provides various interfaces such as USB, RS-232, Ethernet, and analog signals to facilitate integration into OEM systems. The power supply has two DC outputs, and the stable emission current regulation circuit controls the filaments power supply, providing precise and stable current to the X-ray tube. In addition, the power supply also has a comprehensive fault diagnosis circuit, arc monitoring, arc extinguishing, and arc counting functions.

Typical Application:Non-destructive testing,X-ray scanning,Secure applications,Medical applications,Al Visual Recognition

 

Reference Back to list

Features

  • 160kV 225kV 320kV and 450kV Mode
  • 6kW
  • Build in PFC circuit
  • Integrated double filament power supply
  • Solid encapsulation
  • USB Ethernet and RS232 ports

CONTACT US

Address:16 Renxian Street, High-tech Industrial Zone, Dalian, China
Telephone:+86411-84754722
Mobile:+8615642371743
Wechat:15642371743
Email:chengxingqi@teslamanhv.com
URL:www.teslamanhv.com
DaLian Teslaman XingQiCheng

Product Specifications


Input Voltage

208 or 400VAC±10%, three phase, 47-63 Hertz, passive power factor corrected

Input Current

<25 amps per phase for 208VAC. <15 amps per phase for 400VAC.

Output Voltage

Accuracy

0.25%

Stability

≤ 0.1% per 8 hours, after 1 hour warm up

Load

±0.1% of rated output voltage for a full load change

Line

±0.1% of rated output voltage over specified input voltage range

Temperature Coefficient

50ppm/°C

Emission Current

Temperature coefficient

100ppm/°C

Load

±0.1% of rated output current for a change from 30% to 100% of rated output voltage

Line

±0.1% of rated output current over specified input voltage range

Filament

Output

0-6 amps at a compliance of 10Vdc, maximum

Dual Focal Spot

Small and large, selectable via interface signal

Configuration

DC filament drive. Closed loop emission control regulates filament setting to provide desired X-Ray tube emission current

Polarity

Specify negative or positive when ordering

Environment

Work

0°C To +50°C.

Storage

-40°C To +85°C.

Mains Input Connector

MS3106A24-11S

Interface Connector

Digital—Ethernet, RS-232 and USB Analog—25 pin connector

Cooling

Forced air