TIE0001 High precision high voltage divider Accessories

TIE0001 High precision high voltage divider

TIE0001 high-precision high-voltage voltage divider is mainly used for high-voltage measurement under ultra-low ripple and high stability requirements. It adopts potential shielding voltage division methods such as resistance capacity, with high input impedance and good linearity. It adopts special shielding technology to reduce the impact of high voltage on the display value, so as to achieve high stability and high linearity. 

 

Reference Back to list

Features

  • Input: AC220V±10%
  • 50/60Hz.
  • Voltage Level: DC:100kV
  • Stability: 5ppm
  • Accuracy: DC:0.5%
  • Temperature:-25〜40°C
  • Humidity: <85%
  • Separation ratio:10000:1

CONTACT US

Address:16 Renxian Street, High-tech Industrial Zone, Dalian, China
Telephone:+86411-84754722
Mobile:+8615642371743
Wechat:15642371743
Email:chengxingqi@teslamanhv.com
URL:www.teslamanhv.com
DaLian Teslaman XingQiCheng

Product Specifications


Usage Method

1

Securely ground the grounding end of the voltage

divider.

2

Connect the high-voltage wire above the average

pressure ball and tighten the high-voltage wire

with a wiring screw bolt.

3

Insert the high-frequency plug on the instrument

base into the digital display file. The

corresponding test can be carried out at this point.

4

After the test is completed, cut off the high

pressure and make sure that there is no high

pressure before entering the site.

5

When measuring DC, if there is a filter capacitor,

the filter capacitor should be fully discharged so

that the reading indicated by the digital display

meter is "0000" before the relevant operation can

be done.

6

In use and maintenance, attention should be paid

to the cleaning of the surface of the voltage

divider to prevent filth. This instrument should be

placed in a dry place and properly kept.

Precautions

1

In use, 100KThere must be no debris within 2

meters around the V divider. Keep a long

distance between the instrument and the divider

and strictly maintain a safe operating distance.

2

Adhere to the principle of holding and placing

gently, and do not collide, squeeze, etc.

Accessory Part

1RootCoaxial cable test line, 1 grounding wire.