Noise Control of Low-Ripple High-Voltage Supplies in Precision Impedance Analysis and Capacitance Metrology Instruments
An impedance analyzer measuring capacitance to 0.01 percent at 1 kHz uses a 200 V bias supply to polarize the device under test, and the bias noise sets the measurement floor. At the 10 picofarad range, 200 microvolts of bias noise produces a 0.006 percent capacitance error, which is more than half of the instrument's error budget. Noise control therefore drives the entire supply architecture.
The first specification written was spectral, not just total: noise below 20 microvolts RMS in the 0.1 Hz to 10 Hz band, below 60 microvolts in the 10 Hz to 1 kHz band, and below 100 microvolts above 1 kHz up to 100 kHz. A single total-noise number was abandoned after an early prototype met 150 microvolts total but concentrated 90 microvolts in the 1 to 10 Hz band, which happened to sit exactly on the measurement integration window.
Switching noise was removed by using a linear regulator as the final stage. A preregulator supplies 240 V from a 100 kHz converter, and a series pass bank with 12 high-voltage MOSFETs in a cascode arrangement drops the remaining 40 V. The cascode was needed because a single device could not hold 200 V with the required 5 ppm linearity; the cascode splits the voltage and each device sees 18 V.
Reference noise dominates at low frequency. A 200 V output from a 2.5 V reference means a gain of 80, so 1 microvolt of reference noise becomes 80 microvolts at the output. The reference is a buried zener filtered by a two pole RC network with a 0.3 Hz corner, and the measured 0.1 to 10 Hz noise fell from 45 to 14 microvolts after the filter was added. A heated reference was considered and rejected because the thermal cycling during instrument warm-up produced a longer settling time.
Layout and shielding provided the rest of the reduction. The divider uses 20 series resistors in a guard ring driven by a buffer at the same potential, which removes the leakage-induced noise that appears as a 0.2 Hz wander. The output cable is a 1 m coaxial assembly with a driven guard, and cable capacitance of 60 pF is isolated from the regulator by a 100 kohm resistor inside the feedback loop.
Grounding follows a single-reference scheme. The measurement circuit and the bias supply share a star point at the instrument chassis, and any additional connection produces a 30 microvolt loop at 50 Hz that appears as a fixed offset in the capacitance reading. A customer installation with a grounded test fixture introduced exactly that, and the fix was an isolation transformer on the fixture, not a change in the supply.
Calibration stability over time was measured across 14 instruments over 12 months. Output drift was 12 ppm per year on average, with two outliers at 40 ppm. The outliers both had divider modules from a lot with a 35 ppm per degree coefficient instead of 15, and the lot was traced and replaced under warranty. The instrument specification now requires divider lot qualification on every incoming batch.
Thermal design keeps the linear stage inside a safe envelope. The pass bank dissipates 8 W at 200 V and 40 mA, and the heatsink holds the junction at 78 degrees in a 35 degree ambient. Above 40 degrees ambient the supply derates to 25 mA. A fan was avoided because fan vibration produced 15 microvolts of microphonic noise at 120 Hz, which is worse than the thermal margin such a fan would buy.
Protection is kept simple to avoid adding noise. Overvoltage trips at 215 V through a comparator on the divider tap, overcurrent limits at 50 mA by reducing gate drive rather than by shutting down, and a slow-start ramp of 300 ms prevents a capacitor charging transient from tripping the current limit. An active crowbar was evaluated and rejected because the trigger circuit's own noise coupled into the output.
Acceptance testing uses a spectrum analyzer with a low-noise preamplifier and a battery-powered test load. The supply is measured at 25, 100, and 200 V, at 0, 25, and 45 degrees, and at three load currents. Criteria are the three band limits, drift below 20 ppm per hour after a 60 minute warm-up, and no line-frequency component above 5 microvolts. The last criterion is the one most builds fail on first pass.
Practical constraints are documented. The linear stage efficiency is 34 percent at full output, and the instrument draws 60 W from the mains for a 200 V, 40 mA bias, which is acceptable for a bench instrument but rules out battery operation. The 0.3 Hz reference filter also gives a 4 s settling time after a setpoint change, so automated test sequences must allow for that delay.