160 kV Fast-Ramp High-Voltage Technology for Non-Destructive Testing
Industrial X-ray inspection of thick steel castings and pipeline welds requires 160 kV tubes capable of ramping from 10 kV to full voltage in under 1.8 seconds to minimize thermal stress on ceramic inserts while enabling rapid switching between low-kV defect sensitivity and high-kV penetration modes on the same part. Fast-ramp supplies must therefore deliver controlled dV/dt up to 140 kV/s with overshoot below 1.2 % and filament current synchronized within 220 ms.
The generator uses a 42 kHz series-resonant inverter feeding a parallel-fed Cockcroft-Walton multiplier with only six stages to minimize stored energy (11 J total) and intrinsic capacitance. Ramp control is performed by a digitally shaped sinusoidal reference that drives the inverter frequency from 48 kHz at low voltage to 38 kHz at 160 kV, producing an inherently smooth S-curve acceleration profile with peak dV/dt occurring around 88 kV where tube impedance is most favorable.
Overshoot suppression employs an active damping loop that monitors output voltage second derivative via a compensated capacitive pickup and momentarily reduces inverter drive when curvature exceeds a pre-programmed threshold. Resulting voltage settles to ±0.6 % of final value within 180 ms of ramp completion, eliminating the 4–7 kV ringing that previously required 8-second stabilization delays.
Filament synchronization is achieved by pre-heating the cathode to 92 % of operating temperature at 10 kV, then boosting current in a single exponential step timed to coincide with the voltage ramp reaching 72 kV. Emission rise time matches tube voltage increase, preventing cold-cathode sputtering at low kV and space-charge-limited blur at high kV.
Arc handling during ramp uses a two-tier system: minor arcs below 2 J are quenched by instantaneous inverter shutdown for 11 µs followed by seamless resumption at the exact voltage where interruption occurred; major arcs trigger full collapse to 8 kV for 220 ms before restarting the ramp from that point. Average ramp time remains under 2.1 seconds even with three minor arcs per cycle.
Thermal management of the multiplier tank uses forced-oil circulation with variable-speed pump referenced to oil temperature rise rate, maintaining ceramic insulator gradients below 2.8 °C/cm during continuous 60-second-on/60-second-off cycling. Tube focal spot stability supports 0.4 mm resolution at 160 kV through welds up to 110 mm steel equivalent.
These fast-ramp systems routinely achieve full 160 kV in 1.62–1.78 seconds with emission readiness at 99.7 % of steady-state value, enabling inspection throughput increases of 68 % on automated casting lines while extending tube life by 42 % through reduced thermal shock.