Application of Pulse Output Characteristics of Ampoule Inspection High Voltage Power Supplies in Micro-Defect Identification of Glass Containers
A sealed glass ampoule with a pinhole a few micrometres across will pass a visual inspection and fail in storage. The practical test for such a defect applies a high voltage across the wall and looks for the discharge that a pinhole supports, and the supply's pulse shape determines whether the discharge is distinguishable from the corona that every ampoule produces at the tip.
The voltage has to be high enough to break down the gas in a pinhole and low enough not to break down the glass. The two thresholds are separated by a factor of two or three for a thin wall, and the margin is what makes the test possible. A supply that overshoots the setting on the leading edge eats the margin, and the overshoot is the first thing to measure when the false reject rate rises.
Pulse width matters because the discharge through a pinhole takes time to develop. A pulse of a few microseconds may end before the discharge is established, and the test then misses defects that a longer pulse would find. A pulse of a few tens of microseconds is the usual compromise, and the compromise is between sensitivity and the heating of the product.
The corona at the ampoule's tip is the interference that the test has to reject. Every ampoule has a sealed tip with a curvature that concentrates the field, and the corona current there is larger than the pinhole current for a small defect. Separating the two requires looking at the discharge's signature rather than at the total current, and the signature includes the rise time and the pulse's position relative to the applied pulse.
Electrode geometry sets the field distribution and therefore the background. A pair of rollers that contact the ampoule over a defined length produces a uniform field along the body, and a single point contact produces a field that varies by an order of magnitude along the body. The roller arrangement costs more to build and produces a background that is stable enough to set a threshold.
Product conductivity changes the load. A liquid product with a high conductivity shunts the field through the glass wall and reduces the voltage across the wall's defect, and a product with a low conductivity lets the field concentrate in the wall. The same supply setting produces different sensitivities for different fills, and the setting has to be part of the product's recipe.
Crosstalk between adjacent test stations is a real constraint on a filling line. Stations that are close together couple through the product and through the conveyor, and a discharge at one station appears as a signal at the next. Separating the stations by a distance that is large compared with the electrode spacing, and sequencing the pulses so that adjacent stations are not firing together, are the two practical measures.
The reject decision has to tolerate a defined number of events. A single discharge within a pulse is not a defect; several discharges within one pulse, at the same position, are. Counting events per pulse and comparing the count with a threshold is more robust than looking for a single event, because the count averages the random nature of the discharge.
Speed on a filling line is set by the inspection time per ampoule, and the inspection time is set by the pulse repetition. A line that runs at several hundred ampoules per minute gives a few milliseconds per ampoule, which is enough for one pulse and not enough for a train of pulses. A single pulse of the right shape is the design that fits the line.
The high voltage path on a filling line is exposed to cleaning agents and to product spillage, and the insulation degrades faster than in a laboratory. A tracking path across a contaminated insulator produces a discharge that looks like a defect at every ampoule, and the resulting false reject rate is the symptom that leads to the cleaning schedule.
Humidity around the test station changes the corona threshold. A station in a wash down area sees humidity that varies with the line's cleaning cycle, and the corona current varies with the humidity. A threshold that is fixed for dry air rejects good ampoules after a wash, and a threshold that follows the humidity measurement is the practical answer.
Traceability requires that every reject be recorded with the pulse's signature. A record that stores the waveform for a reject allows the cause to be identified later, and a record that stores only the decision leaves the cause unknown. Storage of the waveforms for a shift is a small amount of data and settles most disputes.
The ampoule's wall thickness varies within the tolerance of the forming process, and the breakdown voltage varies with the thickness. A test that is set for the thinnest wall in the batch will not break down a pinhole in the thickest wall. Measuring the thickness optically at the test station and adjusting the pulse's amplitude accordingly is the way to keep the sensitivity uniform.
Commissioning the inspection station ends with a seeded defect trial. Ampoules with known pinholes of three sizes are run through the line at production speed, and the detection rate for each size is recorded. The trial establishes the smallest detectable defect, which is the number the quality specification should quote, rather than the voltage the supply can produce.
The pulse's leading edge is where most of the information lives and where most of the design effort goes. A slow edge ramps the field through the corona threshold gradually and produces a corona that is indistinguishable from the pinhole discharge; a fast edge produces a corona that is sharp and a pinhole discharge that is sharper. Edge rates above a few kilovolts per microsecond separate the two signatures, and the rate is the specification that matters more than the amplitude.
A supply that produces a fast edge pays for the fast edge in radiated interference. The edge couples into the reject signal path and into the line's control network, and the coupling produces false rejects that appear only when the line's other equipment is running. Shielding the signal path and routing the high voltage return as a coaxial structure is the practical fix, and the fix belongs in the mechanical design rather than in the electronics.
Pulse repetition on a line with several stations produces a beat that shows up as a periodic variation in the false reject rate. Two stations firing at slightly different rates produce interference at the difference frequency, and the interference is strongest at the ampoules that pass between the stations. Synchronizing all stations to a common clock removes the beat at the cost of a distributed timing signal.
The product's fill level changes the capacitance of the ampoule and therefore the load the pulse sees. A half filled ampoule has a different capacitance from a full one, and the supply's pulse shape changes with the load. A supply that is specified into a fixed capacitance produces a different amplitude on a partially filled ampoule, and the sensitivity varies with the fill level in a way that the recipe has to compensate.
Defect recognition by waveform is a classification problem with a small number of classes. Corona, pinhole, contamination, and a bad contact each produce a distinguishable waveform, and a classifier trained on a few hundred labelled waveforms is more reliable than a threshold on a single feature. The training data come from the seeded defect trial and from the rejects that the line produces in normal operation.
A reject that occurs at a defined position on the ampoule is a mechanical defect, and a reject that occurs at random positions is a surface defect. Recording the position along with the waveform separates the two, and the separation changes the maintenance action from cleaning the electrode to adjusting the roller pressure.
The electrode's contact pressure affects the field and the background. A roller that presses too hard deforms the ampoule and changes the wall's thickness at the contact, and a roller that presses too lightly loses contact and produces an intermittent open circuit. Pressure that is set by a spring with a defined rate is more repeatable than pressure set by an operator's feel.
Long term drift in the inspection station is dominated by the electrode's wear. A roller's surface becomes polished with use, the contact area grows, and the field distribution changes. Measuring the background corona current at a fixed voltage once per shift tracks the wear, and the replacement interval can be set from the trend rather than from a calendar.
The station's enclosure has to contain the high voltage and let the ampoules pass. An enclosure with a grounded mesh and a defined opening size keeps hands out and lets the field through, and the mesh's aperture has to be small compared with the wavelength of the interference that the pulses produce. A mesh that is too coarse radiates the pulses into the line's control cabinet.
Validation of the inspection station for a new product starts with the product's conductivity and the fill level, and ends with the seeded defect trial at the line's speed. A trial run at a lower speed proves nothing about the production case, because the pulse timing relative to the ampoule's motion changes with the speed.
A pinhole discharge is a small event. The background is what hides the discharge.